FDA Brief

Company record

IsoTis OrthoBiologics, Inc.

1 FDA warning letter since 2023 and 1 recall in the last 12 months.

Get an email the day FDA posts anything new about IsoTis OrthoBiologics, Inc..

Warning letters

  1. IsoTis OrthoBiologics, Inc.

    IsoTis OrthoBiologics in Irvine, CA failed to properly fix and assess possible particulate contamination in its Accell bone filler devices, FDA says.

    High risk

Recalls

ReportedClassFirmProduct and reason
Aug 5 Class II IsoTis OrthoBiologics, Inc.
Device
OsteoBallast, 45 x 17.5 mm (SeaSpine Brand), REF: 56800095. Ballast, 45 x 17.5 mm (IsoTis Brand), REF: 02-8000-095
Demineralized bone matrix (DBM) in resorbable mesh indicated for use with autograft as a bone graft extender in the posterolateral spine and pelvis, were filled with approx. 4.5 cc DBM (fill for 45 ¿ 11 mm configuration) instead of approx. 7.5 cc DBM (fill for 45 ¿ 17.5 mm configuration), which may lead to delayed fusion, incomplete fusion, pseudarthrosis, or lack of bone ingrowth.